TY - JOUR ID - TI - The Response of nuclear track detector (CR-39) to alpha particles in the energy range ( 1.5 – 5.37)MeV استجابة كاشف الأثر النووي CR-39 لطاقة جسيمات ألفا ضمن المدى 5.37)MeV-1.5) AU - Ahmad A. Ibraheem1 , Saja. R. Mohammed2 أحمد عبد إبراهيم العبيدي1 ، سجى رافع محمد الطائي2 PY - 2015 VL - 10 IS - 4 SP - 93 EP - 104 JO - Kirkuk Journal of Science مجلة كركوك للعلوم SN - 30054788 30054796 AB - The aim of this project is to study the effect of the variation of the energy of α - particles on the plastic detector CR-39 in terms of their effect on the diameters of the tracks and their numbers . The plastic detector CR-39 is irradiated by ( 1.5 – 5.37) MeV α-particle emitted from Amretioum Am241 source . Potassium Hydroxide solution (KOH) with concentration 6.25N , at a temperature of 70 oC was used for etching for different etching times . The experimental results had indicated that maximum absorption of the enrgy of α-particals which corresponds to the maximum track diameter is at an energy of 1.75 MeV , and that the maximum number of tracks which is 524 occurs at energy of 2MeV

الهدف من دراسة هو دراسة تأثير تغير طاقة جسيمات الفا على الكاشف البلاستيكي CR-39 من حيث تأثيرها على أقطار الآثار وأعدادها، حيث تم تشعيع الكاشف CR-39 بجسيمات الفا المنبعثة من مصدر الامريشيوم Am 241 واستخدم محلول هيدروكسيد البوتاسيوم KOH بتركيز 6.25N ودرجة حرارة 70oC ولأزمان قشط مختلفة لإظهار آثار جسيمات الفا ولمدى من الطاقات تتراوح بين (1.5 – 5.37) MeV .لقد أظهرت النتائج العملية إن أعلى امتصاص لطاقة جسيمة الفا والذي يقابله أقصى قيمة لقطر الأثر عند الطاقة 1.75 MeV كما إن عدد الآثار تصل أقصى قيمة لها عند الطاقة 2 MeV حيث تساوي 524 أثرا . ER -