TY - JOUR ID - TI - DESIGN AND CONSTRUCTION OF A DIAGNOSTIC PULSED X-RAY TUBE AU - Mohamed Ibrahim PY - 2010 VL - 13 IS - 3 SP - 51 EP - 57 JO - Al-Nahrain Journal of Science مجلة النهرين للعلوم SN - 26635453 26635461 AB - A pulsed X-ray Tube based on the field emission cold cathode principle has been designed andconstructed. The device produced X- ray pulses of duration of (1.2) μ sec and intensities of (90) mRat a distance of (10) cm from the window, when a (150) KV is discharged across the electrodes ofthe tube. The intensity of the x ray radiation at various distances was determined at various cathodeanode spacing to determine its optimum value. Special efforts were made in the design to bedemountable, easy to maintain, and offering a practical possibility for future industrial and medicalapplication

في هذا البحث تم تصميم وبناء انبوب اشعة سينية بنيتعلى تقنيةالانبعاث بواسطة الكاثود البارد. ان الجهاز اعطىاشعة سينية بزمن نبضة قدرها ( 1.2 ) مايكروثانية وشدةاشعاع قدره ( 90 ) ملى رنتكن على مسافة 10 سم منالانبوب عند تسليط 150 كيلو فولت عبر الاقطاب. انالخصائص الاخرى كشدة اشعاع على مسافات مختلفة منالانبوب وافضل مسافة بين الكاثود والانود والتي تعطياعلى شدة تم دراستها ايضا. ان الانبوب صمم ليكون منالنوع القابل للتفكيك، سهل الصيانة مع امكانية استخدامهمستقبلا للاغراض الصناعية والطبية ER -