TY - JOUR ID - TI - تصميم منظومة حقن جرياني لتقدير الفينول AU - احمد صالح فرهود AU - داخل ناصر طه PY - 2012 VL - 20 IS - 2 SP - 618 EP - 632 JO - Journal of University of Babylon مجلة جامعة بابل SN - 19920652 23128135 AB - Flow injection designation have been done with simple procedure and low cost for phenol determination . The estimation condition flow rate , concentration of K3[Fe(CN)6] and H3BO3 as a carrier , reagent volume , reagent concentration , sample (phenol) volume and length of reaction coil were determined . The flow rate was 4.32 ml/min , the carrier concentration(1000,146 )ppm of boric acid and potassium frricyanide respectively ,the reagent concentration was 20 ppm of 4-aminoantipyrine with optimum volume equal 157 µl , sample(phenol) volume was 157 µl and coil reaction length was 150 cm . Under these conditions the interval samples were 120 sample per hour with detection limit 1×10-4 ppm and linearity within (0.01-10)ppm concentration range was 0.998 and RSD equal 0.355 for ten read at 7 ppm of concentration and finally with dispersion coefficient was 1.83.

تم تصميم منظومة حقن جرياني بسيطة العمل قليلة الكلفة لتقدير الفينول . ثم تم دراسة تاثير سرعة جريان التيار الناقل , تركيز مكونات التيار الناقل (حامض البورك وسيانيد البوتاسيوم الحديديكي ) , حجم الكاشف , تركيزالكاشف , حجم النموذج (الفينول) , طول ملف التفاعل . وجد ان الظروف المثلى لتقدير الفينول بهذه المنظومة هي سرعة جريان 4.32 مل بالدقيقة ,وتركيز1000 و146 ppm لحامض البوريك وسيانيد البوتاسيوم الحديديكي على التوالي كتيار ناقل و 20 ppm تركيز كاشف 4- امينوانتي بايرين بحجم 157 µL و حجم 157 µL لنموذج الفينول و 150 سم طول ملف التفاعل . تحت هذه الظروف كان عدد النماذج المحللة 120 نموذج بالساعة وبحد كشف 1×10- 4 ppm وبمدى تركيز 0.01 – 10 ppm وبخطية (R2 ) 0.998 و R.S.D مقداره 0.355 لعشر قراءات عند تركيز 7 ppm ومعامل تشتت مقداره 1.83 . ER -